Ultrasonic milk analyser

The operator has to put a glass filled with water and to press the button Enter
in order to continue the procedure. Then the following message appears on
the display:

Cleaning

Please wait

Do not miss to clean with water because the residues from the alkaline
cleaning solution could react with the acidic solution, which will lead to gas
and sediment formation.

After rinsing with water, the display shows the following:

Put filled with

Detergent glass

and press Enter

to Continue


The operator has to pour alkaline solution in the glass and to press the button
Enter, in order to continue the cleaning.
Then the following message appears on the display:

Cleaning

Please wait


Follow repeatedly sucking/pouring out the cleaning solution and heating it up.
After the cleaning is finished, the cleaning solution is emitted in the reservoir
with the waste liquids.

For maximum cleaning effect it is recommended the cleaning solution to be
preliminary heated up to 40-50 degrees centigrade.