LABEC M-MIA-A User Manual

Page 85

Advertising
background image

Ultrasonic milk analyser

2.5. Electrolyte Replacement (for refillable electrode only).

The reference electrolyte needs to be refilled when the electrode has been
used for a long period, or when the internal electrolyte has dried up. To
accomplish this, follow the procedure described below.

- Remove the protective rubber cap or sleeve;
- Remove the protective rubber sleeve to expose the filling port of the

electrode;

- Remove the old reference electrolyte with a syringe;
- Fill the new reference electrolyte.

2.6. New electrolyte preparation:

- Open the KCl container;
- Add in de-ionized water until it reaches the level of 20 ml;
- Close the container and shake it to dissolve the KCl;
- Add in fresh electrolyte until it reaches the level of the refilling port. The

reference electrolyte used should be 3M(Mol) KCl;

- Replace the rubber sleeve.

2.7. Re-use the electrode.

- Rinse the liquid junction with de-ionized water.

If these steps fail to restore normal electrode response, you may attempt to
rejuvenate it (See: Electrode Rejuvenation).

2.8. Electrode cleaning

Electrodes which are mechanically intact can often be restored to normal
performance by one or combination of the following procedures.

- Salt deposits:
Dissolve the deposit by immersing the electrode in tap water for ten to fifteen
minutes. Then thoroughly rinse with de-ionized water. Wash the electrode pH
bulb in a little detergent and water. Rinse electrode tip in with de-ionized
water.

- Oil/Grease films:
Wash electrode pH bulb in a little detergent and water. Rinse electrode tip
with de-ionized water.

- Clogged Reference Junction:
pH electrodes have junction, which allows the internal fill solution of the
measuring electrode to leak out into the solution being measured. The

Operation manual

MRC.VER.01-02.10

Advertising