LABEC M-MIA-A User Manual

Page 35

Advertising
background image

Ultrasonic milk analyser

If after the second correction the results are unsatisfactory we recommend
making a calibration of the analyser. In dependence of the conditions and
your requirements you may make the calibration using a personal computer
type IBM PC and the company's calibration program or autonomous - by
recalibration.

When making corrections or calibrations be 100% sure in the accuracy of the
reference methods result.

6.4. Recalibrating the milk analyser

6.4.1. Running the analyser in mode Recalibrate

6.4.1.1. Press the button Enter and without releasing it switch on the power
supply of the device, wait for the starting identification messages and release
the button after the following message appears on the display:

Release button

to start setup


After releasing the button on the display is shown:

MA Setup


Followed by the possible to be entered by the operator menus:

Special modes
Corrections
Settings
Tests
------------------------------
pH & Co Meter
Exit


6.4.1.2. By using buttons “up” ▲ and “down” ▼ position on Settings and
press button Enter.

Operation manual

MRC.VER.01-02.10

Advertising