LABEC M-MIA-A User Manual
Page 53
 
Ultrasonic milk analyser
1.To cancel the entered data by pressing the button 0 on the external keypad 
and to start entering them again. 
 
2.To confirm the entered data by pressing the button Enter on the external 
keypad. The display shows currently chosen calibration. The device is ready 
to start measurement. After completing the measurement, the data for the 
deliverer are printed out. 
 
8.5. Measuring high fat samples (option). 
The standard device measures samples up to 25% fat. 
On customer’s request, the device could be produced with possibility to 
measure samples up to 50% fat. The customer can choose which calibration 
to have this possibility and which not, as well as during the process of 
exploitation to change the measuring mode i.e. to pass from measuring 
normal fat percentage towards high and vice versa. 
What the operator sees during these passes is the difference in the speed of 
sucking the sample. For that purpose, the high-fat sample has to be 
preliminary heated up to 30С +- 3С. 
To choose the mode, follow the sequence below: 
 
Setup->Settings->Settings 
Page2->Option 
Select->SelPumpSpeed->Speed for Cal x 
 
After which the display shows: 
 
 Calibr x 
 Pump Speed 
 
Normal OK HiFat
 
By pressing the correspondent buttons the operator can choose the type of 
measurement and to exit the menu. 
When changing the type of measurement on a calibration is necessary a new 
calibration of the device on the new speed to be done. When calibrating 
measuring high fat sample, before starting the calibration procedure, the 
operator has to choose from the menu: 
 
Setup->Settings->Settings 
Page2->Option 
Select->SelPumpSpeed-> HFSpeed for Cal 
 
By which the device passes in a mode of measuring high fat samples. 
Operation manual