LABEC M-MIA-A User Manual
Page 95
 
Ultrasonic milk analyser
 
Co Pass 1/5=xxxx 
Put new sample 
and press Enter 
to start 
 
 
Where xxxx is the result from the first calibration measurement. The operator 
have to put a new buffer, N.B. do not use one and the same buffer more than 
once! Then start the next measurement. This procedure has to be repeated 5 
times. At the end the following message appears on the display: 
 
 
Cond Meter 
Calibr= xxxx 
Switch Off/On 
 
 
Now the operator has to switch off the power supply of the analyzer. After 
switching it on again, the analyser has to be cleaned again with water, which 
ends the calibration of the conductivity measurement system calibration. 
 
Note: 
Another possibility for calibration of analyzer’s conductivity measurement 
function. 
You need conductivity meter. First measure milk with normal acidity with 
conductivity meter and use it as sample for calibrating the analysers 
conductivity measurement function. 
 
4. Co Meter Test. 
Serves for testing the working mode of the milk's sample conductivity 
measurement system. It is used in the production conditions. After this menu 
is chosen, the analyser executes the procedure for sample's measurement 
and the display shows the data, used for obtaining the samples conductivity. 
 
 
Co Meter Test 
CoADC= xxxx 
Power Off - Stop
 
5. Co Meter En/Dis. 
Enables or disables the conductivity measurement system. The following 
message appears on the display: 
Operation manual
MRC.VER.01-02.10