Ultrasonic milk analyser
Analyzer ready
 
Then the analyser is ready for normal measurement. 
 
4.1.1.2. Manual start 
 
The current cleaning may be completed by manual start of the menu 
Cleaning. It is used before starting the menu Final Clean. It serves for 
cleaning the fats from the measuring tract with alkaline cleaning solution. It is 
started using the menu for choosing the working mode of the analyzer. After 
the measurement is completes, by continuous pressing the button Enter, the 
possible analyser operation modes are shown on the display. 
 
 
Cow 
Sheep 
UHT 
Cleaning 
Final Clean 
 
 
By choosing Cleaning the current cleaning is started. The display shows the 
following: 
 
Auto clean
started!
Put empty glass
 
In this way the current cleaning is started. The procedure and operator’s 
actions are described above. 
 
Do not switch off the device at the end of the working day before the 
automatic cleaning procedure is completed. If it is not automatically cleaned 
and there is not a possibility to wait starting the automatic cleaning, then start 
manually the cleaning procedure with alkaline cleaning solution and if it is 
necessary with acidic cleaning solution as it is described below. 
 
Operation manual
MRC.VER.01-02.10