LABEC M-MIA-A User Manual

Page 94

Advertising
background image

Ultrasonic milk analyser

value (6,5 - 13,00
mS/cm (18°C)

Infections damage the tissue of the udder. This
allows sodium and chlorine ions from the blood to
be released into the milk. The concentration of
ions in the milk is thereby raised, and it can more
easily conduct an electrical current - the
conductivity of the milk increases.


Milk conductivity can be used as tests for degree of water evaporation in
condense milk production.
Milk conductivity change notifies of powder (dry) milk solution rate.

2. Conductivity measurement
Conductivity measurement is additional possibility of the analyser and is
delivered on customers request/

3. Co Meter Calibr
Serves for conductivity measuring system calibration. Clean the analyzer
before starting conductivity measurement. (see p. 4.1). You need a standard
buffer with conductivity 5.02[mS/cm] (you may order it for delivery together
with the analyser), with temperature 18

o

С. After starting this mode, the

analyzer makes preparation for measurement and when it is ready, the
following message is displayed:


Co Meter Cal
Put 5.02 buff
and press Enter
to start


The operator has to put the buffer and to start the measurement. The
following message appears on the display:


Measurement
started
Wait please


The buffers' temperature is indicated during measurement. After finishing the
measurement the following message appears on the display:

Operation manual

MRC.VER.01-02.10