LABEC M-MIA-A User Manual
Page 94
 
Ultrasonic milk analyser
value (6,5 - 13,00 
mS/cm (18°C)
Infections damage the tissue of the udder. This 
allows sodium and chlorine ions from the blood to 
be released into the milk. The concentration of 
ions in the milk is thereby raised, and it can more 
easily conduct an electrical current - the 
conductivity of the milk increases.
 
Milk conductivity can be used as tests for degree of water evaporation in 
condense milk production. 
Milk conductivity change notifies of powder (dry) milk solution rate. 
 
2. Conductivity measurement 
Conductivity measurement is additional possibility of the analyser and is 
delivered on customers request/ 
 
3. Co Meter Calibr 
Serves for conductivity measuring system calibration. Clean the analyzer 
before starting conductivity measurement. (see p. 4.1). You need a standard 
buffer with conductivity 5.02[mS/cm] (you may order it for delivery together 
with the analyser), with temperature 18
o
С. After starting this mode, the
analyzer makes preparation for measurement and when it is ready, the 
following message is displayed: 
 
 
Co Meter Cal 
Put 5.02 buff 
and press Enter 
to start 
 
 
The operator has to put the buffer and to start the measurement. The 
following message appears on the display: 
 
 
Measurement 
started 
Wait please 
 
 
The buffers' temperature is indicated during measurement. After finishing the 
measurement the following message appears on the display: 
 
Operation manual
MRC.VER.01-02.10