LABEC M-MIA-A User Manual
Page 36
 
Ultrasonic milk analyser
6.4.1.3. Analogically, position on Recalibrate and press the button Enter. 
 
6.4.2. Making recalibration 
6.4.2.1. Determining the calibration mode 
After starting Recalibrate, the display shows the following: 
 
 
 Recalibrate 
Calibration 1 
Calibration 2 
Calibration 3 
-----------------------------
Edit samp’s 1 
Edit samp’s 2 
Edit samp’s 3 
Edit FrPoints 
-----------------------------
 Exit 
 
 
Using the buttons “up” ▲ and “down” ▼ position on the corresponding 
calibration (for example if you need to recalibrate cow milk, which is assigned 
in the milk analyser as Calibration 1 - cow) and press the button Enter.The 
following message appears on the display: 
 
Calibration 1
Getting ready
 
6.4.2.2. Entering values for the separate sample parameters 
Wait for the sound signal and the following menu on the display: 
 
Cal1 Samp High 
Fat=f.ff 
SNF=y.yy 
Den=d.dd 
------------------------------
Lac=l.ll 
Sol=s.ss 
Pro=p.pp 
Exit 
 
Operation manual
MRC.VER.01-02.10