LABEC M-MIA-A User Manual
Page 34
 
Ultrasonic milk analyser
 
By using buttons “up”▲ and ”down”▼position on the corresponding 
calibration (for example Correction 1 – cow) and press Enter. 
6.3.3.2. Choosing correction parameter 
After choosing calibration mode the display shows the following: 
 
 
 
 
 
 
 
 
 
 
 
 
Using the buttons “up”▲ and ”down”▼position on the parameter to be 
corrected (for example FAT) and press the button Enter. 
6.3.3.3. Making correction 
After choosing parameter (for example fat) the display shows the following: 
 
FAT Correction
0,05
- OK
+
 
Using the buttons “up”▲ and ”down”▼is possible to increase or decrease the 
value of the measured parameter in the above pointed limits. Leaving this 
mode means saving the correction value and activating it. 
6.3.3.4. Making verification 
After the corrections are made put the milk analyser in working mode and 
make several times measurement of reference samples with known values of 
the corrected parameter. If the difference between the values of the 
parameter from the reference methods and milkanalyser are in the limits for 
the parameter it may be considered that the correction is successfully made. 
If the discrepancy between the measurements from the milk analyser and 
classical methods is bigger than is necessary to make second correction 
according above described way. 
Cal1 Correct’s 
FAT 
SNF 
Density 
------------------------------ 
Lactose 
Salts 
Proteins 
Water 
------------------------------ 
Exit 
Operation manual
MRC.VER.01-02.10