LABEC M-MIA-A User Manual
Page 26
 
Ultrasonic milk analyser
Follows rinsing with water: 
 
The operator has to put a glass with water for rinsing the device. 
 
 
 
Put filled with
Water glass
and press Enter
to Continue
 
After rinsing the display shows the following: 
 
End of cleaning
 
After 2 seconds the following appears on the display: 
 
Analyzer ready
 
After this the analyzer is ready for normal measurement or to be switched off 
from the power supply. 
 
4.2. Preparation of cleaning solution 
4.2.1. Preparation of alkaline cleaning solution 
Preparation of 1 % alkaline solution of Lactodaily for circulation cleaning in 
the milk analyzer: 
1. Take the package 100 g concentrated powder chemical Lactodaily 
2. Carefully cut the upper end, paying attention not to spill it. 
3. In appropriate vessel (for example bucket) pour 1 l water. 
4. Add the powder and then again water up to 10 l. 
Then follow the instruction for milk analyzer cleaning. 
 
4.2.2. Preparation of acidic cleaning solution
Preparation of 1 % acidic solution of Lactoweekly for circulation cleaning in 
the milk analyzer: 
1. Take the package 100 g concentrated powder chemical Lactoweekly 
2. Carefully cut the upper end, paying attention not to spill it.