EXFO IQS-12004B DWDM Passive Component Test System for IQS-500 User Manual
Page 138

Testing Multiple DUTs with the LabVIEW Application
126
IQS-12004B
Calibrating the System Wavelength Response
To perform a wavelength response calibration:
1. If the System menu is grayed out, that means no test has been created
since the application was started. Consequently, no calibration can be
performed until you start a new test. For more information, see Starting
a New Test on page 118.
2. Select Wavelength Response Calibration from the System menu of
the main window.
3. Connect start fiber to the specified switch port and power meter
channel.
4. When you are satisfied with the value displayed in the Power Meter
Monitor box, click Start.
5. Repeat steps 3 and 4 for each power meter channel. The connection
information section provides you with the number of the switch port
and the number of the power meter channel indicating where the
patchcord must be connected.
6. When you are done with wavelength response calibration, simply click
Close to return to the main window.
Note: The zone labeled Instructions guides you through the whole wavelength
response calibration process.