Refilpdl, Reforlmandrel – EXFO IQS-12004B DWDM Passive Component Test System for IQS-500 User Manual
Page 291

COM Objects Reference
DWDM Passive Component Test System
279
Measures Interface—Methods
Measures
RefIlPdl
Description
This command performs an IL/PDL reference measurement.
Syntax
object.RefIlPdl(FireEvent)
Parameters
Byval Boolean FireEvent
Response
None
Possible error ErrInitBefore, ErrCalWrBefore, InProgress, ErrAcqType,
ErrAcqStopped, ErrScanRange, ErrOptimizeBefore,
ErrSourceNotEnabled. For more information, see IQS-12004B
DWDM Passive Component Test System (Standard) COM Error
Messages on page 208.
Note
Use only the IQS-5150 polarization state adjuster.
Measures
RefOrlMandrel
Description
This command performs an ORL reference measurement by
mandreling the launch fiber.
Syntax
object.RefOrlMandrel(FireEvent)
Parameters
Byval Boolean FireEvent
Response
None
Possible error ErrInitBefore, ErrCalOrlRtjBefore, ErrCalOrlMandrelRtjBefore,
ErrInProgress, ErrAcqType, ErrAcqStopped, ErrScanRange,
ErrSourceNotEnabled. For more information, see IQS-12004B
DWDM Passive Component Test System (Standard) COM Error
Messages on page 208.