SCPI Commands Reference
238
IQS-12004B
Specific Commands
CONFiguration:KeepIL
Description This command keeps insertion loss curves for all states of
polarization (4), when a PDL acquisition has been performed.
Syntax
CONF:KIL<wsp><KeepIL>
Parameter
The <KeepIL> parameter is a boolean value indicating if the IL
curve is kept for all SOP:
“1” - Keep IL for all SOP
“0” - Do not keep IL for all SOP
Response
None
Example
CONF:KIL1
See also
READ:SOP, READ:CURV:ILPD?
Note
Default: 0
This command requires a lot of memory space as there are four
more curves by channel.
CONFiguration:KeepIL?
Description The response to this query indicates if insertion loss curves are
kept for all states of polarization.
Syntax
CONF:KIL?
Parameter
None
Response
A value representing if the IL is kept for all SOP:
“1” - The IL is kept for all SOP
“0” - The IL is not kept for all SOP
See also
READ:SOP?, READ:CURV:ILPD?