Bwupperbound, Centralwavelength – EXFO IQS-12004B DWDM Passive Component Test System for IQS-500 User Manual
Page 388

COM Objects Reference
376
IQS-12004B
BandPassResult—Properties
BandPassResult
BwUpperBound
Description
This read-only property returns the upper-bound wavelength
value over the bandwidth interval, based on the current power
level (from the peak). The power level is defined in the
PowerLevel property of the parent object. BwUpperBound value
is expressed in nm.
Syntax
object.BwUpperBound
Parameters
None
Response
A Double value
Access
Get
BandPassResult
CentralWavelength
Description
This read-only property returns the wavelength at the midpoint
between the lower and the upper wavelengths based on the
power level of the parent object. The value is expressed in nm.
Syntax
object.CentralWavelength
Parameters
None
Response
A Double value
Note
The midpoint that is returned is calculated for the current
power level (from the peak) as specified in the PowerLevel
property of the parent object.
Access
Get