COM Objects Reference
DWDM Passive Component Test System
275
Measures Interface—Methods
Measures
AcqIlPdlDut
Description
This command performs an IL/PDL measurement on the DUT.
Syntax
object.AcqllPdlDut(FireEvent)
Parameters
Byval Boolean FireEvent
Response
None
Possible error
ErrInitBefore, ErrHardNotCompatible, ErrCalWrBefore,
ErrRefIlPdlBefore, ErrInProgress, ErrAcqType, ErrAcqStopped,
ErrScanRange, ErrSourceNotEnabled. For more information,
see IQS-12004B DWDM Passive Component Test System
(Standard) COM Error Messages on page 208.
Note
Available only with the IQS-5150 polarization state adjuster.
Measures
AcqOrlMandrelDut
Description
This command performs an ORL measurement on the DUT.
Syntax
object.AcqOrlMandrelDut(FireEvent)
Parameters
Byval Boolean FireEvent
Response
None
Possible error
ErrInitBefore, ErrCalOrlRtjBefore, ErrCalOrlMandrelRtjBefore,
ErrRefOrlMandrelBefore, ErrInProgress, ErrAcqType,
ErrAcqStopped, ErrScanRange, ErrSourceNotEnabled. For more
information, see IQS-12004B DWDM Passive Component Test
System (Standard) COM Error Messages on page 208.