Section 3 – Installation

Dukane Manual Part No. 403-574-01

CAUTION

The power cord is equipped
with a three-prong, ground-
ed-type plug for your safety.
Whenever a two-slot re-
ceptacle is encountered,
we strongly recommend
that it is replaced with
a p r o p e r l y g r o u n d e d
three - lead rec eptacle.

Have a qualified electri-
cian do the replacement in
accordance with the Na-
tional Electrical Code and
local codes and ordinances.
DO NOT cut off the power
cord grounding prong or
alter the plug in any way.

Table 3 - II

Standard IEC AC Power Cord Part Numbers

Connecting Cables

(Quick Start Guide)

Details about the various system connectors and their pin
assignments are covered in Section 4.

Manually Operated Probe System

(Hand Probe)

Step 1. Ground the generator chassis using the supplied

14-Gauge wire, and attach it to the grounding stud:

A

in Figure 3 -6.

Step 2. Attach the hand probe’s HD–15 system input con-

nector to J2 on the I/O panel:

B

in Figure 3 -6.

Step 3. Attach high–voltage coaxial cable to J1, the ultra-

sound output connector:

C

in Figure 3 -6.

Step 4. Connect the AC power cord to the IEC power inlet

connector on the ultrasonic generator (

D

in Fig-

ure 3 -6), and plug the other end into an approved
AC outlet.

Power Cords

The AC line cords supplied with the standard generators
are matched to the ultrasonic output power rating and the
continent of specified use. See Table 3-II.

Continent of Use

Power Cord

Part Number

Power

North America

200 - 1110

240V, 15A

200 - 1541

240V, 10A

Continental Europe

200 - 1111

240V, 16A

200 - 1542

240V, 10A

NOTE

Refer to Section 10, Options for informa-
tion about optional features.