System test, Gh e, Caution – Dukane ULTRASONIC GENERATOR/POWER SUPPLY LS 403-574-01 User Manual

Page 64: Iq series, Of the system power output level bar graph

Advertising
background image

Page  5 

iQ Series

,

Ultrasonic Generator/Power Supply LS User’s Manual

Dukane Manual Part No. 403-574-01

System Test

To test the system’s ultrasound operation, perform the
following steps.

Step 1. For this test, the generator must be ONLINE. If

the OFFLINE

LED

is showing yellow (D in Fig-

ure 7-3) press the ON/OFFLINE key.

The ONLINE

LED

(G in Figure 7-5) should turn

green.

Step 2. Connect a probe to the generator ultrasonic out-

put.

Step 3. Place the probe so that the tip of the horn is not

in contact with anything. If you are using a hand
probe, you may feel a slight vibration or sensation
in your hand. This is normal. There should not be
any loud or unusual noise.

Step 4. Momentarily press the TEST control key (E in

Figure 7-5).

Step 5. The first segment

LED

of the SYSTEM POWER

OUTPUT LEVEL

bar graph (

H

in Figure 7-5)

turns green. The probe will operate as long as the

TEST

key is pressed.

Step 6. When the TEST key is released, the SYSTEM

POWER OUTPUT LEVEL

bar graph turns off

(gray). This test cannot be performed with the

generator OFFLINE.

CAUTION

Any unusually loud noise
from the probe stack in-
dicates that it has been
improperly assembled.
Check the probe stack
fo r c o r re c t as s em -
bly and proper torque.

LOAD

80

60

100

OVER

40

0

20

SYSTEM POWER OUTPUT LEVEL

SYSTEM OPERATING MODE

LOAD

80

60

100

OVER

40

0

20

SYSTEM POWER OUTPUT LEVEL

SYSTEM OPERATING MODE

ON

LINE

OFF

LINE

POWER

TEST

# 1 PART COUNT 125
WELD TIME 0.080 S
POWER 1050 W
ENERGY 350 J

ON

LINE

# 1 PART COUNT 125
WELD TIME 0.080 S
POWER 1050 W
ENERGY 350 J

INF

O

O

N

/O

FF

LIN

E

TE

ST

C

AN

CE

L

ENT

ER

G

H

E

Figure 7 - 5 System Test