AAMI/IEC Standards

The AAMI and IEC standards specify sound levels, CO

2

concentration, and

thermal characteristics for incubators. The standards are used by
manufacturers when designing and manufacturing incubators. INCU was
designed with consideration of the standards, and can perform testing to
satisfy many testing requirements.

Incubator Standards

Standard#

Description

IEC 601-2-19

Infant Incubator Tester Safety Requirements

IEC 601-2-20

Safety Requirements for Transport Incubators

IEC 601-2-21

Infant Radiant Warmer Standard

ANSI/AAMI 1136-1997

Infant Incubators

INCU Applications

The following section provides information describing how INCU

can be

used to perform testing in consideration of the standards.
1.

Temperature – Standard:

During steady state condition, the

INCUBATOR TEMPERATURE shall not differ from the AVERAGE
INCUBATOR TEMPERATURE by more than 0.5

°C (1°C transportable)

during at least 1 hour at the control temperatures of 32°C and 36°C.

INCU: The user shall check the oscillation from min to max in steady
state.

2.

Temperature - Standard:

The AVERAGE TEMPERATURE in each

point A,B,C,D,E, shall not differ from the AVERAGE INCUBATOR
TEMPERATURE (test at a set T of 32

°C - 36°C) by more than ±0.8°C

(

±1.5°C transportable).

INCU: In any position of the tilted mattress, it shall not differ by more
than

±

1.0

°

C (

±

2.0

°

C transportable). Calculate manually the difference

between the mean value at the center and the other sensors readings.

3.

Temperature - Standard:

The TEMPERATURE as measured by the

skin temperature sensor shall not differ from a reference sensor
temperature by more than 0.3

°C in steady temperature condition.

INCU: Check the difference between the value given by the T2 sensor
and the value displayed by the skin sensor when placed in close
proximity to the T2 sensor.

Note: