Incubator Analyzer

INCU Setup

3-7

Air Flow Sensor: The

Air Flow Sensor is detachable; however, it must be

inserted to perform measurements. An error "fault sensor" will be displaced
if not inserted or if the Hot Wire is broken in the sensor. The INCU will not
perform airflow measurements in this condition.

The Air Flow Sensor (Figure 3-5) rotates 360 degrees. For best results,
position the Air Flow Sensor perpendicular to the airflow within the
incubator. Incubators utilize different airflow patterns; thus, the user needs to
be familiar with the particular airflow pattern utilized by the incubator under
test to properly position the Air Flow Sensor.

Warning! Do not use in the presence of oxygen!

The Air

Flow Sensor should

not

be used in an oxygen-enriched

environment. The Air Flow Sensor utilizes a hot-wire technique,
which can be a source of ignition.

Figure 3-5. Placement of Air Flow Sensor