Fluke Biomedical INCU Incubator Analyzer User Manual

Page 41

Advertising
background image

Incubator Analyzer

Data Acquisition

4-3

6. Once the PC program is up and running, select the

Connect

command and then

Config

.

The

Config

command is used to select the measurement interval and

total recording period, and to document the serial number of the
incubator under test as shown in Figure 4-2.

Figure 4-2. Using the Config menu

7. To select the measurement interval, click on the scrolling arrows

located next to

Sampling interval

. You can choose from 1 minute

up to 10 minutes in one-minute intervals.

The measurement interval specifies how often the sensors are
sampled for data. That is, if the measurement interval is one minute,
all seven INCU sensors are then polled every minute and the results
are recorded.

8. To select the recording period, click on the scrolling arrows located

next to

Total measurement time.

You can choose up to 35 hours for

the recording period (with a 10-minute sampling period).

The recording period specifies the duration of the test. That is, if the
measurement interval is one minute and the recording period is set
for one hour, the test will last for one hour, and during that time, 60
measurements for each sensor will be recorded. The maximum
measurement capacity of the INCU is 1470 (210 records x 7
parameters).

9. Enter the serial number for the incubator under test in the field next

to

Incubator serial no

.

Note:

You

must

enter a serial number in this field to proceed.

10. From the

File

menu, choose

Print

to print out a

monitoring sheet

(see Figure B-1 in Appendix B, Report Examples). This sheet
contains the INCU configuration information along with fields for
manufacturer, type, department, etc. Use the

monitoring sheet

to

record events that occur during the incubator-testing phase.