INCU
Operators Manual

iv

Hazard Warnings


☛

Warning! Power Rating. The INCU mains power input must be connected using an

external power supply that provides voltage and current within the specified rating for
the system.

☛

Warning! Internal Voltage. Always turn off the power switch and unplug the power

cord before cleaning the outer surface of the INCU.

☛

Warning!

Liquids. Avoid spilling liquids on the analyzer; fluid seepage into internal

components creates a potential shock hazard. Do not operate the instrument if
internal components are exposed to fluid.

☛

Warning! “Do not use in the presence of oxygen”. The Air Flow sensor is hot wire

technology and is a source of combustion if used in the presence of oxygen within
the incubator. Use the INCU only in ambient oxygen conditions.


☛

Warning! “Use only approved battery charger." The INCU contains a lead-acid

rechargeable battery. Use only the approved charger with proper voltage and current
ratings; otherwise, damage to the unit may result.


Precautions

The following precautions are provided to help you avoid damaging the system:

☛

Caution: Service. Authorized service personnel should service the INCU. Only

qualified technical personnel should perform troubleshooting and service procedures
on internal components.

☛

Caution: Environmental Conditions. Do not expose the system to temperature

extremes. Ambient temperatures should remain between 18°C to 40°C. System
performance may be adversely affected if temperatures fluctuate above or below this
range.

☛

Caution: Do NOT Immerse. Clean only with a mild detergent, and gently wipe down

with a clean, lint-free cloth.


☛

Caution: Electromagnetic radiation may affect the noise measurement.