Fluke Biomedical INCU Incubator Analyzer User Manual
Page 61

Incubator Analyzer
Performance
Specifications
A-3
Noise
Measuring Range:
30- 80 dbA
Resolution: 0.1
dbA
Accuracy:
±
5 dbA @ 30-80 dbA
Display
2 x 16 Super Twist LCD
Measuring Interval
Via PC:
Adjustable from 1 to 10 minutes
Internal Memory
Capacity:
1470 measurements
(210 records of 7 parameters)
Supply
Power Requirements:
Maximum over-voltage:
264 VAC
Input voltage range:
100 to 240 VAC
Input frequency range:
47 to 63 Hz
Power consumption:
< 60 Volt Amperes
Fuse rating:
2A Slow Blow
Battery:
Rechargeable sealed lead-acid type
NP7-6 YUASA, 6V, 7 Ah
dimensions 151 x 34.x 101.0 mm
Operates for 24 hours continuously without the
low-battery warning
Charger 600 mA minimum, 8 VAC-DC
(Charger must be agency-approved, depending
on the country)