Key details – Fluke Biomedical INCU Incubator Analyzer User Manual
Page 29

Incubator Analyzer
INCU Keypad Functions
3-3
Key Details
event selec t
20.4 C Temp. 1
1:210 Evt. 0
º
Figure 3-2. INCU keypad display
3. The sensor number that is displayed flashes
to indicate that the measured values have
been recorded.
4. Press this key a second time to pause (to
allow time for adjustment, for example).
During this pause phase, the channel
number stops flashing.
5. Press this key a third time to resume
recording. There is no limit to the number
of interruptions possible.
Start/Pause Key
1. Press this key for approximately 1 second to
begin recording. This function is only active
once the INCU has been connected to a PC and
a measurement cycle has been initialized.
2. A second display line appears (see Figure 3-2)
with the following information:
• Measurements that have already elapsed
• Total measurements
• The number of events recorded
(maximum 5).