Performance specifications – Fluke Biomedical INCU Incubator Analyzer User Manual

Page 60

Advertising
background image

INCU
Operators Manual

A-2

Performance Specifications

This appendix provides the device specifications for the INCU Incubator
Analyzer. For more information, please contact your Fluke Biomedical
Service representative.

Temperature Measurements

Conduction:

1 sensor in contact with mattress
(bottom of unit)

Convection
(Inside the incubator):

3 sensors (T1, T3, T4)

Radiation:

1 sensor (T2) connected to the so-called “radiant
baby”

Convection
(Inside or outside the
incubator):

1 sensor T2

Measuring Range:

5

°C to 70°C

Resolution:

0.1

°C

Accuracy:

Reading Value

±0.5°C ± 1 LSB for the range

from 25

°C to 40°C

Storage Temperature

-20°C to 50°C

Operating Temperature:

10°C to 40°C

Dimensions

:

27 x 14 x 20 cm

Relative Humidity

Measuring Range:

0% RH to100% RH

Resolution:

0.1% RH

Accuracy:

±5% RH (for range 0 to 90% RH
@ 25

°C to 40°C) or,

± 5.3 % RH (for range 0 to 100% RH
@ 25

°C to 40°C)

Humidity must be without condensation.

Air Flow

Measuring Range:

0.1-0.7 m/s

Resolution: 0.01

m/s

Accuracy:

From 0.1 m/s to 0.5 m/s reading
±0.1 m/s @ temperature 25°C to 40°C
and humidity 50% RH

± 15% RH