T1, T2, and T3 Convection Sensors:

Figure 3-4 shows the recommended

placement for temperature sensors T1-T3 for closed incubators utilizing
forced convection. With the sensors mounted to the provided clips,
temperature accuracy, response, and uniformity can be monitored throughout
the area occupied by the infant at the height of approximately 10 cm above
the mattress as specified by standards. In this case, T2 is being utilized as a
convection temperature sensor.

Humidity Sensor (Fixed):

The humidity sensor is fixed and is located on the

inside of the right-hand cover as shown in Figure 3-4.