Air Flow Sensor

Plug the Air Flow Sensor into the jack on top of the INCU where it is labeled
"Air Flow." Ensure that the Air Flow Sensor is installed before attempting to
use the INCU. Otherwise, a “Sensor Fault” condition will result, indicating
the absence of this sensor. To bypass the “ Fault sensor : 6” message, press
any key. INCU operation will continue with normal operation with the
exception of recording Air Flow parameters. Results will be displayed as
9.99 m/s.

The Air Flow Sensor is calibrated for each INCU unit. If the Air Flow Sensor
is lost or damaged, the unit must be returned for service. See "Support" in
Chapter 5, Safety, Displayed Messages, Troubleshooting, and Support.

Radiant Baby Assembly

Figure 2-1 shows the radiant baby assembly.

Figure 2-1. Radiant baby assembly

Warning! Do not use in the presence of oxygen!

The Air Flow Sensor

uses a hot-wire technology that can be a source of ignition if
used in oxygen-enriched environments. INCU should not be
used in conditions where oxygen levels are greater than
ambient.