Displayed messages – Fluke Biomedical INCU Incubator Analyzer User Manual

Page 55

Advertising
background image

Incubator Analyzer

Troubleshooting

5-3

Displayed Messages

Possible INCU displayed messages are described below.

Displayed Message

Description

Self Testing

INCU is performing internal self-checks to
determine proper operating condition before taking
measurements. Any errors encountered during this
test will be displayed on Line 1. Refer to Chapter 5
for an explanation of possible error conditions from
the INCU and possible cause and resolution.

Low Battery

The battery should be charged without delay.
When the battery voltage is less than 5.35 volts
DC, the message will be displayed. The battery
voltage is displayed on power-up during the self-
test routine.

When the message appears, use the AC Charger
for normal operation and/or to recharge the internal
battery. When the internal battery voltage is greater
than 5.85 volts DC, the message will not be
present.

Note: Regular charging of the battery increases its
life considerably. The battery must never be
completely discharged.

Note: When this message appears, no data
transfer between the INCU and the PC is possible.
Use the AC charger. When the message
disappears, data transfer is possible.

End of Measurements

Indicates that the data acquisition is complete.
Connect the INCU to the PC to transfer data
between the INCU and the PC.

Note: The recorded measurements can be
retrieved using the PC even if the cycle has been
interrupted before this message is displayed.

Data in Memory

The data transfer has not been completed. Using
the PC Software, transfer the data to the PC. Data
must be transferred to the PC in order to take
further readings in memory.

Ready

Displayed after a new configuration has been sent
from the PC software and stored in the INCU unit.

Fault Sensor

There is an error with the subject sensor. Refer to
Chapter 5 to determine corrective action for each
respective sensor error.