Fluke Biomedical INCU Incubator Analyzer User Manual

Page 57

Advertising
background image

Incubator Analyzer

Support

5-5

Troubleshooting (Cont.)

Problem

Possible Cause

Corrective Action

Fault sensor: 6

Air Flow Sensor not
plugged in.

Plug in Air Flow Sensor, or
press any key to bypass
the error condition. The
INCU will perform all other
measurements with the
exception of Air Flow. All
Air Flow readings will be
displayed and stored as
9.99 m/s.

Hot Wire broken in Air Flow
Sensor

Replace Air Flow Sensor; unit
must be sent back to Factory
for recalibration.

INCU did not record all
the measurements that it
was configured for.

If the low battery
message appeared while
in the middle of data
collection, then the INCU
will stop the recording at
that time.

Recharge the battery, or
use the AC Charger. All
data is stored in the INCU
up to the point of the low
battery message. Transfer
the results to the PC using
the PC Software program.

The

stop key was

pressed prior to
completion of the desired
recording cycle.

All data is stored in the
INCU up to the point of the
low battery message.
Transfer the results to the
PC using the PC Software
program.

Advertising