Specifications, General, Output power measurement – Fluke Biomedical QED 6 User Manual

Page 22

Advertising
background image

QED 6

Users Guide

1-10

Specifications

The following are specifications for the Analyzer. Please contact your Fluke
Biomedical service representative for more information regarding the device
specifications.

General

Display

.....................................................................2-line x 24-character LCD

supertwist alphanumeric

Power

......................................................................One 9 V Alkaline (Duracell MN1604 or

equivalent); 12 hours continuous
operation; low battery indication; 9 V
battery eliminator input.

Weight

....................................................................4.54 lb

Dimensions

.............................................................26.67 x 24.13 x 10.16 cm

Environmental Operating Specs

Storage Temperature ..........................................-25 to 50 °C
Operating Temperature .......................................0 to 40 °C
Maximum Humidity..............................................90 % Relative Humidity

Output Power Measurement

Load Resistance......................................................50

Ω

±1 % non-inductance

(< 10 µH), 160 W

Range

1000 J..................................................................0-1000.0 J
100 J....................................................................0-100.0 J

Resolution................................................................0.1 J
Max. Vage

1000 J..................................................................5500 V
100 J....................................................................1750 V

Max. Current

1000 J..................................................................110 A
100 J....................................................................35 A

Measurement...........................................................1000 J: 66 ±5 V
Trip Levels ...............................................................100 J: 20 ±5 V
Pulse Width .............................................................1-50 ms
Accuracy

1000 J Range ......................................................±2 % of reading
100-1000 J .......................................................... ±2 Js
100 J Range ........................................................±2 % of reading

±0.1 J