The LCD displays the reading for about two seconds.

Generating Test Waveforms

The Analyzer generates a series of test waveforms designed to verify the
accuracy of ECG machine / monitors. These waveforms, shown in Table 2-2 ,
are available for simulation via ECG jacks or paddle plates and are calibrated
for lead II at 1 mV.

Table 2-2. Test Waveforms

ECG Performance Arrhythmia

30 BPM

30 BPM Pulse

Atrial Fibrillation

60 BPM

60 BPM Pulse

Atrial Flutter

120 BPM

2 Hz Triangle (2 mV)

Atrial Tachy

180 BPM

0.125 Hz Square (50 % dc)

Idioventricular

240 BPM

2.0 Hz Square (50 % dc)

PVC

10 Hz Sine

R on T

40 Hz Sine

Run

50 Hz Sine

Ventricular Fib.

60 Hz Sine

Ventricular Tachy

100 Hz Sine

To generate waveforms for testing:

1. Turn the defibrillator to be tested to

ON

.

2. Connect the ECG patient leads to the Analyzer as shown in Figure 2-7.