Generating test waveforms – Fluke Biomedical QED 6 User Manual
Page 44

QED 6
Users Guide
2-18
fcf020.eps
Note
The LCD displays the reading for about two seconds.
Generating Test Waveforms
The Analyzer generates a series of test waveforms designed to verify the
accuracy of ECG machine / monitors. These waveforms, shown in Table 2-2 ,
are available for simulation via ECG jacks or paddle plates and are calibrated
for lead II at 1 mV.
Table 2-2. Test Waveforms
ECG Performance Arrhythmia
30 BPM
30 BPM Pulse
Atrial Fibrillation
60 BPM
60 BPM Pulse
Atrial Flutter
120 BPM
2 Hz Triangle (2 mV)
Atrial Tachy
180 BPM
0.125 Hz Square (50 % dc)
Idioventricular
240 BPM
2.0 Hz Square (50 % dc)
PVC
10 Hz Sine
R on T
40 Hz Sine
Run
50 Hz Sine
Ventricular Fib.
60 Hz Sine
Ventricular Tachy
100 Hz Sine
To generate waveforms for testing:
1. Turn the defibrillator to be tested to
ON
.
2. Connect the ECG patient leads to the Analyzer as shown in Figure 2-7.