Testing high level out, Measuring peak voltage, current, and overshoot – Fluke Biomedical QED 6 User Manual

Page 46

Advertising
background image

QED 6

Users Guide

2-20

Testing High Level Out

All waveforms available through the ECG jacks are simultaneously output
through the High Level jacks. This scheme offers the user a 1-volt peak signal
for testing purposes.

To test a High Level signal, use an oscilloscope and a scope probe to measure
the output waveform on the high level output. Refer to Generating Test
Waveforms
, above, for procedure.

Note

Use a scope probe to guarantee signal integrity.

Measuring Peak Voltage, Current, and
Overshoot

To measure the peak voltage and current of the defibrillator pulse:

1. Power up the defibrillator to be tested and select the energy output,

following the manufacturer’s instructions.

2. Power up the Analyzer.

Main Menu 1

displays:

fcf002.eps

3. Press the

PEAK

softkey to access the

--- V --- A

(current and voltage)

menu:

fcf025.eps

Pressing the

Peak / OVER

softkey toggles the measurement between

Peak

current and voltage and

Over

current and voltage.