Measuring charge time (models m and h) – Fluke Biomedical QED 6 User Manual

Page 47

Advertising
background image

Operation

Measuring Charge Time (Models M and H)

2

2-21

4. Set the range switch appropriately, as follows:

Select the 1000 joule high range for defibrillator outputs over 100

joules or for an unknown defibrillator output power.

Select the 100 joule low range for outputs under 100 joules.

5. Simultaneously press the two defibrillator paddles onto the contact

electrode plates on the front of the Analyzer.

6. Initiate a discharge from the defibrillator.

7. Observe the LCD on the Analyzer and record the defibrillator voltage and

current.

Note

The LCD continues to display the reading until the next defibrillator
pulse is fired.

Measuring Charge Time (Models M and H)

To measure the charge time of Models M and H:

1. From

Main Menu 1

, press

more

to access

Main Menu 2

. The following

displays:

fcf026.eps

2. From

Main Menu 2

, press the

CHRG

softkey to display the following:

fcf027.eps

3. Set the range switch appropriately, as follows:

Select the 1000 joule high range for defibrillator outputs over 100

joules or for an unknown defibrillator output power.

Select the 100 joule low range for outputs under 100 joules.