Fluke Biomedical QED 6 User Manual

Page 36

Advertising
background image

QED 6

Users Guide

2-10

fcf006.eps

4. Press the

WAV>

softkey to browse through a list of available waveforms

in Energy Mode. These waveforms are described in Table 2-1.

Table 2-1. Available Waveforms

Waveform Description

ECG90

The default waveform; after a discharge, the ECG90 resumes.

VFIB Ventricular

Fibrillation

VTACH

125 BPM, VTACH

VTAC2

240 BPM, Monomorphic

VTAC3

300 BPM, Polymorphic

ASYS Asystole

The waveform appears through the ECG adapters and Analyzer front
panel paddles and is available to trigger an automatic defibrillator to
discharge.

After the defibrillator discharges, the output switches to a

90 BPM ECG

waveform. If the range switch is set to

Pacer

, the following message

appears momentarily:

fcf007.eps

4. Set the range switch appropriately, as follows:

Select the 1000 joule high range (

1000 J

) for defibrillator outputs

over 100 joules or for an unknown defibrillator output power.

Select the 100 joule low range (

100 J

) for outputs under 100 joules.

5. Simultaneously press the two defibrillator paddles onto the contact

electrode plates on the front of the Analyzer. See Figure 2-5.

Advertising