Pacemaker refractory period testing – Fluke Biomedical QED 6 User Manual

Page 50

Advertising
background image

QED 6

Users Guide

2-24

Note

Pacer voltage and current are displayed as average voltage and
current. If a printer is connected to the Analyzer, the printout also
documents peak voltage and current. If computer control is being
used, no peak values are available. All voltage measurements are
referenced to the internal 50

Ω

load.

Pacemaker Refractory Period Testing

To test the pacer refractory period:

1. Set the pacer in demand mode.

2. From

Main Menu 1

, press

more

to access

Main Menu 2

. The following

displays:

fcf026.eps

3. Set the range switch to

pacer

. Otherwise the unit beeps and displays the

following message:

fcf029.eps

4. From

Main Menu 2

, press the

PACE

softkey to display the

Pacer

Tests

menu:

fcf032.eps

5. Press

RFP

to select refractory period testing. Refer to Figure 2-9 for

setup. The following menu appears.