Operation

Measuring

Synchronization

2

2-17

Sync time measurements are performed as shown in Figure 2-8, below:

R-

W

a

v

e Time

R

QRS COMPLEX

90 BPM

N

SR

Q

Defi

b

P

u

lse

Peak

Q-

W

a

v

e Time

DEFIB PULSE

fcf019.eps

Figure 2-8. Sync Time Measurements

5. Set the range switch appropriately, as follows:

Select the 1000 joule high range for defibrillator outputs over 100

joules or for an unknown defibrillator output power.

Select the 100 joule low range for outputs under 100 joules.

6. Place the defibrillator in synchronous mode.

7. Simultaneously press both defibrillator paddles to the contact plates of the

Analyzer.

8. Initiate a discharge from the defibrillator.

9. Press the

ENRG

softkey to view the energy readings on the display, as

shown below.