Pacemaker (non-invasive) testing – Fluke Biomedical QED 6 User Manual

Page 48

Advertising
background image

QED 6

Users Guide

2-22

4. Press the two defibrillator paddles onto the contact electrode plates on the

front of the Analyzer.

5. Press the

START

softkey and initiate the defibrillator charge cycle.

6. As soon as the defibrillator reaches full charge, discharge it, noting the

time (in seconds) on the display. The maximum for the Analyzer is 60
seconds. After 60 seconds, the Analyzer displays

OVER

.

Pacemaker (Non-Invasive) Testing

1. From

Main Menu 1

, press

more

to access

Main Menu 2

. The following

displays:

fcf026.eps

2. Set the range switch to

pacer

. Otherwise the unit displays the following

message:

fcf029.eps

3. From

Main Menu 2

, press the

PACE

softkey to display the

Pacer

Tests

menu:

fcf030.eps

4. Connect the output from the pacer to the Analyzer, as shown in Figure

2-9.