Programming an automatic test sequence – Fluke Biomedical QED 6 User Manual
Page 52
Advertising

QED 6
Users Guide
2-26
Programming an Automatic Test Sequence
The Model H can store in memory up to 28 automatic sequences to fully test
defibrillator performance according to protocol. Standard defaults for
Programs 0-27 for the Model H are listed in Table 2-3.
Table 2-3. Autosequencing Defaults (Model H)
Performance Parameter
Default
Print Heading
Yes
Energy Measurements
10 J
100 J
200 J
300 J
360 J
Energy Limits + / -
5%
Charge Time
Yes
Sync Time
Yes
Peak No
Overshoot No
ECG performance
30 BPM
120 BPM
240 BPM
2 Hz Sq
.125 Hz Sq
10 Hz Sin
40 Hz Sin
50 Hz Sin
60 Hz Sin
100 Hz Sin
2 Hz Tri
60 BPM PLS
Fib
Pacer No