Programming an automatic test sequence – Fluke Biomedical QED 6 User Manual

Page 52

Advertising
background image

QED 6

Users Guide

2-26

Programming an Automatic Test Sequence

The Model H can store in memory up to 28 automatic sequences to fully test
defibrillator performance according to protocol. Standard defaults for
Programs 0-27 for the Model H are listed in Table 2-3.

Table 2-3. Autosequencing Defaults (Model H)

Performance Parameter

Default

Print Heading

Yes

Energy Measurements

10 J

100 J
200 J
300 J
360 J

Energy Limits + / -

5%

Charge Time

Yes

Sync Time

Yes

Peak No

Overshoot No

ECG performance

30 BPM

120 BPM
240 BPM

2 Hz Sq

.125 Hz Sq

10 Hz Sin
40 Hz Sin
50 Hz Sin
60 Hz Sin

100 Hz Sin

2 Hz Tri

60 BPM PLS

Fib

Pacer No