Operation

Running an Automatic Test Sequence

2

2-31

Note

The changes are saved until the program is modified again, or the
Analyzer is reset to factory defaults.

Running an Automatic Test Sequence

To run an automatic test sequence:

1. From

Main Menu 2

, press the

AUTO

softkey to display the

Autosequences

menu:

fcf046.eps

2. Press the

RUN

softkey to select an autosequence program, which are

numbered 0 - 27.

fcf047.eps

3. Press the

+

or

-

softkey to increase / decrease the program numbers.

4. When the desired program number displays, press the

START

softkey to

start the selected program.

The Analyzer prompts the user through the complete autosequence
program. Data output (if requested) to a printer or computer occurs after
the test sequence has been run.

Note

If the range switch is not already set to the appropriate range or to

pacer

for PACE tests, the Analyzer sounds an audible alarm, and a

warning message appears until the condition is corrected.