Fluke Biomedical QED 6 User Manual
Page 39

Operation
Evaluating Ability to Fire
2
2-13
OFF
DEFIB
Defibrillator/Pacer
PACER
QED-6
DEFIBRILLA
TOR A
NA
LYZER
STERNUM (-)
BAT
TE
RY / 9VD
C
RS-232 / 10101
ECG
V
C
LL
F
LA
L
RL
N
RA
R
DEFIB
APEX (+)
PA
CE
R
1000 j
100 j
SOFTKEYS
fcf010.eps
Figure 2-7. ECG Lead Configuration
3. From the
Energy
menu, press the
WAV>
softkey until the
VFIB
option
appears; then press the
VFIB
softkey.
A ventricular fibrillation waveform is simulated by the Analyzer through
the ECG jacks and paddle plates. When a discharge is complete, the
Analyzer outputs a 90 BPM Normal Sinus Rhythm.
4. Press the
WAV>
softkey until the
VTACH
option appears; then press the
VTACH
softkey.
A ventricular tachycardia is simulated by the Analyzer through the ECG
jacks and paddle plates. When the discharge is complete, the Analyzer
outputs a 90 BPM Normal Sinus Rhythm.