Defibrillator pulse playback, Viewing oscilloscope output, Preparation for viewing – Fluke Biomedical QED 6 User Manual

Page 40

Advertising
background image

QED 6

Users Guide

2-14

Defibrillator Pulse Playback

The Analyzer allows the user to play the defibrillator pulse waveform for the
purpose of analysis. Playback is accomplished using a strip recorder or
defibrillator monitor through the ECG jacks or scope output. The waveform
can also be reviewed on an oscilloscope through the high-level ECG outputs.

To play back the defibrillator pulse:

1. Power up the defibrillator and the Analyzer.

2. From

Main Menu 1

, press the

ENRG

softkey to access the

Energy

menu, and press the

WAV>

softkey to access the

WAV

submenu.

3. Connect the ECG patient leads to the Analyzer as shown in Figure 2-7.

4. After the defibrillator discharges, press the

PLAY

softkey. The last

defibrillator pulse is replayed.

Viewing Oscilloscope Output

The Analyzer provides two banana jacks for real-time viewing on an
oscilloscope with storage capability

To view the output on an oscilloscope, carry out the steps under each of the
subheadings, below:

Preparation for Viewing

1. Connect the Oscilloscope to the Analyzer, using a banana plug and a

scope probe to ensure signal integrity, as indicated:

Ground – Attach the ground from the scope probe to the common

(black) jack on the Analyzer.

Output – Attach the positive lead of the scope probe to the

defibrillator output.

Advertising