Header – Fluke Biomedical QED 6 User Manual

Page 59

Advertising
background image

Operation

Printing the Analyzer Report Header

2

2-33

Fluke Biomedical
Control #:
Serial #:
Model #:
Mfr.
Location:
Technician :
Date:

Settin

g

Actual

---------

20.0

J

Qwave Sync Time

43.9 ms

Rwave Sync Time 73.1

ms

Ener

g

y

19.9

J

Over Volta

g

e

0

V

Over

Current

0

A

Ener

g

y:

19.9

J

PACER OUTPUT MEASUREMENTS
Pulse

Rate:

40

PPM

Pulse

Width:

19.7

ms

Peak Current:

1

8

.0 ma

Peak Volta

g

e: 0.9

V

Ave

Current

0.0

ma

Ave Volta

g

e:

0.0

V

PACER REFRACTORY MEASUREMENTS
Selected Pacin

g

Rate:

40 PPM

Pulsed Refractory Period: 329.2 ms
Sensed Refractory Period: 179.2 ms

Char

g

e

Time: 05

sec

Ener

g

y:

19.

8

J

Header

fcf050.eps

Figure 2-10. Manual Output with Header