Defib waveform playback, Arrhythmias, Scope outputs – Fluke Biomedical QED 6 User Manual

Page 24

Advertising
background image

QED 6

Users Guide

1-12

Defib Waveform Playback

Time Base Expansion .............................................100:1 @ 25 mm/sec paper speed;

each division equals 40 ms

Amplitude Scaling

Lead II (RA-LL)

1000 J Range ..................................................1 mV = 3000 V
100 J Range ....................................................1 mV = 900 V

ECG Output
1000 J Range ..........................................................0.5 V = 3000 V
100 J Range ............................................................0.5 V = 900 V

Arrhythmias

Afib, Vfib, Atach, Vtach, Aflutter, RUN, PVC, R on T, Idioventricular
Rate Accuracy .........................................................±1 %
Amplitude.................................................................Fixed at 1 mV Lead II

(RA-LL)
Fixed at 1.1 mV (Apex-Sternum)

Amplitude Accuracy.................................................±2 (RA-LL) ±10 % (Apex-Sternum)

Scope Outputs

ECG Hi-Level...........................................................Fixed at 1 V

Accuracy: ±2 %

Defib Output ............................................................Real Time
Pacer Range............................................................1 V = 3.11 V
1000 J Range ..........................................................1 V = 1450 V
100 J Range ............................................................1 V = 440 V
Amplitude Accuracy.................................................±2 % of scale