Thorax

109

Hints

• You could repeat the same protocol by simply click-

ing the chronicle with the right mouse button for
“repeat“, e.g., when both non-contrast and contrast
studies are required.

• For lung cancer evaluation, this protocol can be com-

bined with the protocol “Neck Routine”.

• Low dose lung images are usually evaluated using

lung window setting. Soft tissue/bone window set-
tings may be used to visualize the presence of calci-
fications in the nodules.

• It is essential to use the same protocol for follow-up

studies to check for progression.

LungLowDose

kV

130

Effective mAs

30

Rotation time

1.0 sec

Slice collimation

5.0 mm

Slice width

8.0 mm

Feed/Rotation

10.0 mm

Pitch Factor

1.0

Increment

8.0 mm

Kernel

B70s

CTDI

Vol

3.23 mGy

Effective dose

Male: 1.69 mSv

Female: 2.08 mSv

Contrast medium IV injection

Start delay

30 sec.

Flow rate

2.5 ml/sec.

Total amount

50 – 70 ml