Innerear, Children innerear – Siemens syngo CT 2005C User Manual

Page 200

Advertising
background image

200

Children

InnerEar

Indications:

High Resolution Spiral mode for inner ear studies, e.g.,
malformations of the inner ear, inflammatory
changes, pathologies of the mastoid process, tumor
processes of the pyramids, post-traumatic changes,
etc.

A range of 2.5 cm will be covered in 22 sec.

* The conversion factor for a 7-year-old child, and a
scan range of 25 mm was used.

InnerEar

kV

80

Effective mAs

125

Rotation time

1.5 sec

Slice collimation

1.0 mm

Slice width

1.25 mm

Feed/Rotation

2.0 mm

Pitch Factor

1.0

Increment

1.0 mm

Kernel

H80s

CTDI

Vol

8.09 mGy

Effective dose

Male: 0.04 mSv*

Female: 0.04 mSv*

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 200 Friday, April 8, 2005 9:55 AM