* The conversion factor for a 7-year-old child, and a
scan range of 150 mm was used.

Hints

• Children with a body weight of more than 55 kg

should be examined with an adult protocol.

• The first and second recon jobs are defined for visu-

alization of the mediastinum and the lungs, respec-
tively.

• Please change the mAs value according to the

weight of the child:

45 – 54 kg 37 mAs.

ThorRoutine

2

nd

reconstruction

kV

130

Effective mAs

26

Rotation time

1.0 sec

Slice

collimation

2.5 mm

Slice width

5.0 mm

5.0 mm

Feed/Rotation

10.0 mm

Pitch Factor

2.0

Increment

5.0 mm

5.0 mm

Kernel

B41s

B60s

CTDI

Vol

2.80 mGy

Effective dose

Male: 1.46 mSv*

Female: 1.72 mSv*

Contrast medium IV injection

Start delay

exam dependent

Flow rate

dependent upon needle size/
Access site

Total amount

1 – 2 ml per kg of body weight