Scan and Reconstruction

Pitch

In single slice CT:
Pitch = table movement per rotation/slice collimation

E.g.,: slice collimation = 5 mm,
table moves 5 mm per rotation, then pitch = 1.

With the Siemens Multislice CT, we differentiate
between:

Feed/Rotation, the table movement per rotation

Volume Pitch, table movement per rotation/single
slice collimation.

Pitch Factor, table movement per rotation/complete
slice collimation.

E.g., slice collimation = 2 x 5 mm,

table moves 10 mm per rotation,

then Volume Pitch = 2, Pitch Factor = 1.

With the SOMATOM Spirit, the pitch, slice, collimation,
rotation time, and scan range can be adjusted. The
pitch factor can be selected from 0.5-2.

Pitch Models

Pitch 1

Pitch 1.5

Pitch 2