Specials

PolyTrauma

Two times 2 combined recon jobs are predefined, head
with neck and thorax with abdomen.

A range for the Head of 11 cm will be covered in
20 sec., a range for the Neck of 14 cm will be covered
in 30 sec.

Take a new Topogram for the thorax and abdomen
ranges.

Head

Neck

kV

130

130

Effective mAs

110

90

Rotation time

1.5 sec

1.0 sec

Slice collimation

5.0 mm

2.5 mm

Slice width

6.0 mm

3.0 mm

Feed/Rotation

10.0 mm

5.0 mm

Pitch Factor

1.0

1.0

Increment

6.0 mm

3.0 mm

Kernel

H31s

B50s

CTDI

Vol

25.05 mGy

9.70 mGy

Effective dose

Male:

1.02 mSv

Female:

1.08 mSv

Male:

1.69 mSv

Female:

1.81 mSv

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 178 Friday, April 8, 2005 9:55 AM