Osteo, Spine osteo – Siemens syngo CT 2005C User Manual

Page 144

Advertising
background image

144

Spine

Osteo

This is the scan protocol for the syngo Osteo CT appli-
cation package to assist the Physician with the quanti-
tative assessment of vertebral bone mineral density
(BMD) in the diagnosis and follow-up of osteopenia
and osteoporosis.

For obese patients, increase the mAs value to 120.

Load all ranges in the application “syngo Osteo CT”.

For further information, please refer to the chapter
syngo Osteo CT“.

Osteo

Osteo

Osteo

kV

80

80

80

Effective mAs

81

81

81

Rotation time

1.0 sec

1.0 sec

1.0 sec

Slice

collimation

5.0 mm

5.0 mm

5.0 mm

Slice width

10.0 mm

10.0 mm

10.0 mm

Feed/Scan

0.0 mm

0.0 mm

0.0 mm

Kernel

S80s

S80s

S80s

CTDI

Vol

2.23 mGy

2.23 mGy

2.23 mGy

Effective dose

Male:

0.03 mSv

Female:
0.08 mSv

Male:

0.03 mSv

Female:
0.05 mSv

Male:

0.03 mSv

Female:
0.04 mSv

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 144 Friday, April 8, 2005 9:55 AM