Dose information – Siemens syngo CT 2005C User Manual

Page 31

Advertising
background image

Dose Information

31

For spiral scan protocols, the indicated mAs is the
effective mAs per image. The correlation between tube
current mA and effective mAs of spiral scans on a Mul-
tislice CT scanner is given by the following formula:

where Slice collimation refers to the collimation of one
detector row, and nrow is the number of used detector
rows.

Pitch Factor =

Feed/Rot

nrow x Slice collimation

mA = effective mAs x Pitch Factor

Rot Time

Effective mAs = mA x RotTime/Pitch Factor