Headroutine_child, Children headroutine_child – Siemens syngo CT 2005C User Manual

Page 194

Advertising
background image

194

Children

HeadRoutine_Child

Indications:

Routine head spiral studies, e.g., brain tumors, cranial
trauma, hydrocephalus, hemorrhaging and other
abnormalities etc.

A range of 8 cm will be covered in 27 sec.

* The conversion factor for a 7-year-old child, and a
scan range of 80 mm was used.

Head

kV

130

Effective mAs

60

Rotation time

1.5 sec

Slice collimation

2.5 mm

Slice width

3.0 mm

Feed/Rotation

5.0 mm

Pitch Factor

1.0

Increment

3.0 mm

Kernel

C30s

CTDI

Vol

13.66 mGy

Effective dose

Male: 0.55 mSv*

Female: 0.60 mSv*

C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 194 Friday, April 8, 2005 9:55 AM

Advertising